A damp-heat soak sits at 40 °C and 93 %RH. Twenty minutes into the ramp that follows, the chamber air still holds enough water for a dew point near 39 °C. Anything in the working volume that is two d...
A damp-heat soak sits at 40 °C and 93 %RH. Twenty minutes into the ramp that follows, the chamber air still holds enough water for a dew point near 39 °C. Anything in the working volume that is two d...
The performance of these two types of equipment is basically identical, while they differ in control modes. Both are used to pre-treat products and their contents through repeated temperature cycles...
This document outlines the essential handling, transportation, and site placement specifications for industrial temperature and humidity test chambers to ensure operational longevity and measurement...
Most reliability failures in laboratory testing do not come from the product under test. They come from the equipment that is supposed to create the test conditions. A chamber that drifted by 2°C dur...
Environmental Testing / Cold Reliability Design a vehicle-mounted controller that passes every functional bench test at 23 °C, and it can still fail hard at -35 °C. A display driver loses calibration,...
You have a temperature-humidity test chamber set to 25°C and 60% relative humidity. You plan to cool the specimen to 15°C. Before you start the ramp, you need to know whether moisture will condense o...
Imagine a 200-liter environmental test chamber running a humidity profile at 40°C and 90% relative humidity (RH). The controller computes a dew point of 37.9°C. If the measured dew point is off by ju...
Put a batch of lithium-ion cells into a chamber, program 85°C and 85% RH, and walk away for 500 hours. Whether that test produces trustworthy data depends less on the nameplate and more on how precis...
If you are selecting an environmental test chamber, chamber temperature is the specification that determines whether your test gives you useful data or misleading data. A chamber that cannot hold its...
If you have a temperature reading and a dew point reading, you can already compute relative humidity without any additional instrument. The principle is straightforward: the dew point tells you how m...
When you place a lithium-ion cell into a chamber and ramp it toward 150 °C, the reading on the screen can turn from a routine validation curve into a safety decision in less than a minute. Battery sa...
You are standing in front of a stability chamber, and the controller reads 25.0 °C with a dew point of 15.0 °C. The recipe calls for 50% RH at that temperature. Is the chamber performing as expected?...